5. No display after pressing :
Check whether the trigger Mode is “Normal” or “Single”, and if the trigger level exceeds the waveform range. If yes, set the trigger level to the middle or change the
trigger Mode to “Auto”.
Note: press “Auto Scale” could automatically replace the above setting.
6.The waveform displays like ladder:
(1) The horizontal time base may be too low, you can increase it to improve the horizontal resolution so as to make a good waveform displaying.
(2) The lines between the sample points may also cause ladder-like displaying if the “Type” in menu of “DISPLAY” is “Vectors”. Please turn the “Type” to “Dots” to
solve the problem.
7.USB storage can’t be recognized:
(1) Check if the USB ash disk can work normally.
(2) Check if the USB Device Host can work normally.
(3) Make sure the USB disk being used is of ash type, the instrument does not support USB of hardware type.
(4) Restart the instrument and then insert the USB to check it.
(5) If it is still in abnormal use, please contact with SIGLENT.
Trouble
shooting