User Manual

5- SDS1000X-E Series Digital Oscilloscope
Probe Compensation
You should properly compensate the probe at rst use of it. Non-compensated or inadequate compensated probe may cause inaccurate measurement. The
following steps are about probe compensation:
1. Perform step 1, 2 , 3 and 4 of “Functional Inspection”.
2. Check the displayed waveforms and compare them with the following gure.
3. Use a nonmetallic driver to adjust the low-frequency compensation adjustment hole on the probe until the waveform changes to be correct as the gure above.
Quick start
Under
Compensated
Compensated
Correctly
Over
Compensated