CAN Triggering
This part will provide a brief introduction and description for the operation of the LIN
trigger.
To introduce the trigger conditions
Start the oscilloscope will be triggered at the start bit of a frame.
Remote the oscilloscope will be triggered by a remote frame with specified ID.
ID the oscilloscope will be triggered by a remote or data frame that have specified
ID.
ID+DATA the oscilloscope will be triggered by data frame that have specified ID
and data.
Error the oscilloscope will be triggered by an error frame.
Operation steps:
1. Press Setup to enter the TRIGGER function menu.
2. Press Type and select Serial.
3. Press Protocol and select CAN.
4. Press Trigger Setting to enter the CAN TRIG SET menu.
5. Press Condition and select the trigger condition by Universal Knob:
If you select the REMOTE and ID condition:
a. Press ID Bits to set the length of ID (11bits or 29 bits).
b. Press Curr ID Byte and use Universal Knob to select the byte that you want
to set.
c. Press the ID and set the IDs value by Universal Knob.
(Tips: In order to make is convenient for the operator to set the parameters, ID b
is split into several bytes. For example, if the IDs length is 11 bits, it will be split
into two bytes, a byte includes 8 bits. If 1st byte is selected, only the 8 least