To Set the LAN
Do the following steps to set the oscilloscope to communicate with PC via LAN:
1. Connect the oscilloscope to your local area network using the network cable.
2. Press the Utility button on the front panel to enter the UTILITY function menu.
3. Press the IO Set softkey to enter the I/O SET function menu.
4. Press the LAN softkey to enter the LAN setting interface, see the picture below:
Figure 67 LAN Setting Interface
5. Press the F1 softkey (the first softkey below the screen) continually to go to the DHCP
line; then turn the Universal Knob to select Enable or Disable.
Enable: the DHCP server in the current network will assign the network
parameters (such as the IP address) for the oscilloscope.
Disable: you can set the IP address, subnet mask, gate way manually.
Turn the Universal Knob to select the desired value.
Push the Universal Knob to change item horizontally.
Press the F1 softkey to go to the next line.
Press the Save/Recall softkey to save the current setting.
Press the Single button to exit the setting interface.
6. Use EasyScopeX or NI Vista to communicate with the oscilloscope.