SSG5000X User Manual 90
effects accordingly.
Statistics
The default is “off”. When activated, statistics indicate the measured mean,
maximum, minimum and the count of each measurement...
On: Turn on the statistics function. Each statistical parameter is displayed on
the parameter bar. Statistical parameters include average, minimum,
maximum, and statistics. Click the “Clear button to clear all current statistics
and start new statistics.
Off: Turn off statistics. Statistical parameters will be hidden
Auto Zero
When the power meter is zeroed, the zero measurement deviation and noise
influence can be reduced, and the accuracy of the RF power measurement can be
improved.
Disable: The zero buttons will be hidden.
Internal Zeroing: In this way, the power sensor remains connected to the RF
signal during the zeroing process. Zero buttons is displayed.
External Zeroing: When this mode is selected, you must remove the RF signal
to be measured before the power sensor can be zeroed. Otherwise it will affect