SSG5000X User Manual 92
Level Offset
The default is “Off”. Click the slide switch to switch the power offset switch status.
On: The values can be set by the touch screen keyboard or front panel keypad.
The displayed value will be the actual measurement plus the offset. This
feature makes measurements easier when there are amplifiers and
attenuators in the middle of the signal chain.
Off: The display is consistent with the actual reading of the power sensor.
Averaging Mode
The default is “Auto”. Click the slide switch to switch the averaging mode.
Auto: Automatically configure the average number of times based on the
current parameters. Fixed noise parameter settings will be automatically
hidden.
Manual: Manually set the average number times, with the parameter are
behind.
Fixed Noise: The power sensor will control the internal noise that does not
exceed the set value of the fixed noise parameter. Fixed noise parameter
settings will be displayed. It can be set by the touch screen keyboard or the
front panel keypad.
Note: If the power sensor does not have a “fixed noisemode, this option will