User Manual

SIGLENT
SSG5000X User Manual 93
not be displayed in the drop down box.
Logging
The default is “Off”. Click the slide switch to switch the logging.
On: The source will record the measurements and save it in a log file.
Off: The signal source does not record the measurements.
Level Control
As shown in the example, the sensor measures a proportional power in defined
time intervals, derived from a coupler. It considers optionally the given
S-parameters and returns the results to the generator. The signal generator
compares the measured level with the set value and adjusts its output level
accordingly. This allows you to control the external signal level continuously and
reliably reach a constant input level at the DUT in real time.
In practice, an RF splitter is needed to split the RF signal. One of them is connected
to the DUT and the other is connected to the power meter. The signal source is for