SSG5000X User Manual 94
acquisition and compensation. Power compensation can be used to minimize
cable losses, the attenuation of passive networks, and the amplification of the
signal by a power amplifier and the frequency response of each device in the link
as the frequency changes.
Note: When the function is turned on, the frequency or level sweep and the power
sensor measurement function will be turned off, and vice versa.
Sensor State
The default is “off”. Activate/deactivate level measurement by the power sensor.
On: Turn on the power sensor level control function, and the measurement
control refreshes the power sensor measurement value in real time.
Off: Turn off the power sensor measurement function.
Measurement
Indicate the current reading of the sensor. Select the unit used for result display.
You can change the unit for the results display: dBm, dBμV, uV, mV, V, nW, uW, mW,
W.