SSG5000X User Manual 126
The waveform segment selection area lists all waveform segments in the current
volatile memory and their corresponding sampling points. Select a segment to set
the marker points in the right marker setting area.
Marker Setting Area
In the marker setting area, each row of parameters in the table represents the
marker settings of a segment:
First Mkr Point: Set the sample point corresponding to the first marker point
of the segment.
Last Mkr Point: Set the sample point corresponding to the last marker point
of the segment (the last marker point should always be less than or equal to
the number of points in the segment, and greater than or equal to the first
marker point).
Skipped: Set the number of sample points you want to skip, so that only one
sample point will be output every corresponding number of sample points
in the segment.
When marker points are set, they do not replace points that already exist, but
are set in addition to existing points. For example, set the marker point range of
a segment to 1-30, and then add a marker point range of 20-50. The marker
signal consisting of these two marker point ranges starts on sample point 1, and
ends at on sample point 50.