SIGLENT
SSG5000X User Manual 160
2. Set the Frequency & Level
a) Press FREQ, and enter the 1GHz with the numeric keypad, or click the
frequency on the touch screen to pop up the keypad to input 1GHz.
b) Press LEVEL, and enter the 0dBm with the numeric keypad, or click the
amplitude on the touch screen to pop up the keypad to input 0dBm on the
keypad.
3. Set the frequency of flatness calibration
a) Press LEVEL->Flatness-> , select the Fill Type as Manual Step;
b) Set the Start Freq to 4GHz, Stop Freq to 5GHz, Points to 11;
4. Perform amplitude calibration
a) Click Fill Flatness with Sensor button and a Collecting dialog will pop up.
After the data collection is finished, it will automatically return to the flatness list page
and give the corrected amplitude correction value for each frequency.
3.3 Output a modulated signal
In this section, we will show how to enable modulation by configuring the SSG to
output an AM modulated signal with the following characteristics: Carrier