SSG5000X User Manual 162
b) Set the modulation depth to 80%;
c) Set the modulation frequency to be 10 kHz;
d) Set the AM state open;
4. Enable the RF modulation output
a) Press MOD ON/OFF and then press RF ON/OFF. The backlights of MOD ON/OFF
and RF ON/OFF will be illuminated. RF and MOD labels in the status bar will
change from gray to blue.
At this time, the signal on the [RF OUTPUT 50Ω] connector will match the settings
configured above.
3.4 Output a pulse train
A pulse train is a user-defined sequence of high and low RF levels with specific time
durations for each part. In this example, we will show how to configure the RF
source to output the following pulse train:
Serial
number
Positive pulse
width
Negative pulse
width
Repetitions
1
10ms
30ms
1
2
20ms
20ms
2