User Manual

SIGLENT
SSG5000X User Manual 165
related parameters according to actual application requirements.
1. Restore factory settings
a) Press UTILITY or touch Preset Type-> Default.
b) Press PRESET or reset the system settings to perform the reset operation.
2. Set the baseband signal
Press I/Q or select IQ MOD block from the Home page, and then click Custom.
a) Open the Data Source setting interface, and set Data Setup to PN9, Symbol
Rate to 1MHz, and Symbol Length to 512;
b) Open the Modulation setting interface, and set Mod Type to QAM, QAM Type to
32QAM;
c) Open the Filter setting interface, and set Filter Type to RootCosine, Filter Alpha
to 0.35, Filter Length to 128, OverSampling to 2;
d) Set Custom State on;
3. Set the carrier frequency & amplitude
a) Press FREQ, and enter the 1GHz with the numeric keypad, or click the
frequency on the touch screen to pop up the keypad to input 1GHz.
b) Press LEVEL, and enter the -10dBm with the numeric keypad, or click the
amplitude on the touch screen to pop up the keypad to input -10dBm on the