(3) If a key is not working, the numeric keyboard connection might be loose or the
numeric keyboard is broken.
Do not disassemble the instrument by yourself and contact SIGLENT.
3 Correct setting but incorrect waveform output
(1)No RF output
Check that the signal cable is securely connected to the corresponding [RF
OUTPUT 50 Ω] port.
Check the cable for damage.
Check if the RF button backlight is lit. If not lit, press this button to light it and the
user interface status bar RF flag turns blue. The RF output is now turned on
correctly.
Check if the signal output amplitude is too small, and adjust the output
amplitude appropriately.
(2)No modulation on the RF output
Check that the signal cable is securely connected to the corresponding [RF
OUTPUT 50 Ω] port.
Check the cable for damage.
Check that both the MOD and RF button backlights are lit and that the modulation
switch needs to be turned on.