Check if the modulation parameters are appropriate and modulate the
modulation parameters appropriately.
If an external modulation source is used, make sure that the external source is
connected correctly and has an output and should work within the range
established by the source.
4 Sweep is abnormal:
(1)Scanning stagnation
The user interface frequency zone/amplitude zone displays a scan progress bar
indicating that a scan operation is in progress. If there is stagnation, check the
points.
Open at least one scan type: Select Frequency, Amplitude, or Frequency &
Amplitude according to the SWEEP scan type status.
If the single scan mode is enabled, a single scan is performed when the single
software meets the trigger condition.
If the scan trigger mode is not automatic trigger, press SWEEP trigger mode to
automatically trigger to determine if the scan trigger is missing and the scan is
blocked.
If the point trigger mode is not automatic trigger, it is automatically triggered by
the SWEEP trigger mode to determine if the point trigger is missing and the scan is
blocked.