Determine if the dwell time setting is too large or too small for the scan to be
visible.
Make sure at least two points are set in the step scan or list scan.
(2) In the list or step scan, the amplitude does not change
Verify that the scan type is set to Amplitude or Frequency & Amplitude.
If the current scan type is set to Frequency, the amplitude value does not change.
5 USB device cannot be recognized:
(1) Check if the USB device is connected to other instruments or the computer can
work normally.
(2) Confirm the flash type U disk device used. This instrument does not support the
hard disk type U disk device.
(3) After restarting the instrument, insert the USB flash drive to check.
(4) If you are unable to use the USB flash drive normally, please contact.
6 Inaccurate measurement results or insufficient accuracy:
Users can obtain detailed descriptions of technical indicators from the back of this
manual to calculate system errors and check measurement results and accuracy
issues. To achieve the performance metrics listed in this manual, you need:
(1) Check if the RF signal source is within the calibration period (1 year).