User Manual

SIGLENT
SSG5000X User Manual 25
This can be useful if the circuit or device shifts or alters the output frequency .such
as an external mixer, by setting an offset; you can read the mixer signal’s
frequency directly on the RF generator front panel.
1. Keypad operation
Press FREQ->Freq Offset to make the focus fall on this parameter item.
Use the numeric keypad to input the desired value and press the unit button to
select the desired unit. Optional units are GHz, MHz, kHz, and Hz. Press ENTER
to select the current unit.
Press ENTER or press the knob to enter the parameter editing state, and then
use the left and right direction keys to move the cursor to the desired position.
Press the up and down direction keys, rotate the knob or press the numeric
keypad to modify the value. Press ENTER, knob or ESC to exit the editing mode.
2. Touch screen operation
Click SHORTCUT->FREQ-> Freq Offset to enter the interface of frequency offset
setting.
Click the frequency offset parameter, and then pop up the keypad interface,
then click the number symbol in the keypad to change the value. Click the unit
to modify the unit. Click the check mark to confirm the modified value and
selected unit.