SSG5000X User Manual 33
Flatness
The "Flatness" function adjusts the amplitude at specific frequencies by
mathematically adjusting the output by applying user-defined level correction lists.
Correction is performed by the user-defined table values being added to the output
level for the respective RF frequency. With frequencies which are not contained in
the list, the level correction is determined by interpolation of the closest correction
values. If there is only one point, it will not perform correction.
The flatness correction function can adjust the RF output amplitude corresponding
to the frequency point within the frequency range of the instrument to compensate
for losses in the cable, or losses caused by other equipment.
1. Flatness Menu
Keypad operation
Press LEVEL->Flatness to make the focus fall on the flatness menu, then press
the knob or ENTER key to switch flatness status. When opened, the status bar
displays the blue UF letters.
Touch operation
Click on SHORTCUT->LEVEL ->Flatness to make the focus fall on flatness switch.
Then click on the switch button to switch the flatness status.