SSG5000X User Manual 35
5) Save
Press to enter the file save/call page, save the correction data in the list, see
"file management" for details.
6) Return
Press return to the previous menu.
7) Set
Press to enter the set of flatness list filling.
After entered into the set of flatness list filling, you can choose one of the following
three methods to fill flatness list with power sensor.
Flatness List
In the flatness list edit page, insert the frequency list. Then enter the set of
flatness list filling, choose Flatness method, and click Fill Flatness With
Sensor. The amplitude correction values in the flatness list will be generated.
Manual Step
Choose Manual Step method, several hidden options will be enabled. Edit
these options to generate frequency list, and then click Fill Flatness with
Sensor; amplitude correction values in the flatness list will be generated.
Sweep List
Sweep List method will generate frequency list according to the sweep list
you have edited in the RF Sweep function. And then you can click Fill Flatness