SSG5000X User Manual 38
level sweep progress bar.
Step Sweep
The step sweep state is on by default. Click the slide to switch the on-off state. Click
to enter the parameter setting menu of step sweep.
Note: The step sweep and list sweep are mutually exclusive. One will shut off
automatically when the other is on.
1) Start Freq: Initial frequency set point of the sweep.
2) Stop Freq: Final frequency set point of the sweep.
3) Start Level: Initial amplitude level set point of the sweep.
4) Stop Level: Final amplitude level set point of the sweep.
5) Dwell time: The interval time between two sweep points.
6) Sweep points: The number of points in a sweep .Every point is interpolated
between the start parameter and stop parameter.
7) Sweep shape: Click the drop-down box, Sawtooth and Triangle can be selected
Sawtooth: Sweep period is always from the start frequency or start level to the
stop frequency or stop level. The sweep sequence is similar to a "Sawtooth
wave".
Triangle: Sweep period is always from the start frequency or start level to the
stop frequency or stop level, and then return back to the start frequency or
start level. The sweep sequence is similar to a "triangle wave".