SSG5000X User Manual 43
then the device will start one time sweep after receiving a specified polarity TTL
pulse.
Point Trigger
The default point trigger mode is Auto “. The SSG5000X provides four types of
point trigger: “Auto”, “Key”, “Bus” and “Ext”.
Click the drop-down box of point trigger mode to select the desired type:
Auto: The default mode. If the sweep mode is “continuous”, just select one
sweep type to start and sweep through each point continuously in one sweep
cycle. If the sweep mode is “single”, first click Execute single sweep button to
sweep through each point in one sweep cycle.
Key: If the sweep mode is “continuous”, the device will sweep one point at a
time after every press of the Trigger button on the front panel or click Click to
trigger button on the touch screen. If the sweep mode is “single”, you will
need to click Execute single sweep button first, and then press the Trigger
button once for every point in the sweep. The sweep will stop when finish one
sweep cycle.
Bus: If the sweep mode is “continuous”, the device will start a sweep after
you send a “*TRG” command. If the sweep mode is “single”, you will need