User Manual

SIGLENT
SSG5000X User Manual 55
PM Rate
Select MOD ->PM->PM Rate, then you can set the PM rate.
1. The sine wave modulation frequency range is 0.01Hz~100kHz
2. The square wave modulation frequency range is 0.01Hz~20kHz
PM Deviation
The setting range is . For the definition of N, refer to the data
sheet.
When the modulation source is set to Int “, the set value is actually the
frequency offset of the RF output.
When the modulation source is set to Ext”, it is related to the amplitude of
the external input modulation signal. Offset of RF output = amplitude of input
signal × external modulation sensitivity.
When the modulation source selects “Int+Ext”, the set value is the maximum
value of the total offset, the internal source accounts for 50%, and the external
source accounts for 50%. RF output offset = set value × 0.5 + external input signal
amplitude × external modulation sensitivity.