SSG5000X User Manual 58
Pulse Mode, Pulse Period, and Pulse Width and so on.
2. Ext
Use an externally applied modulation signal. The external signal is input via the
[PULSE IN/OUT] connector.
At this time, setting items such as pulse type, pulse period, pulse width, trigger
mode, and pulse output are hidden.
Pulse Polarity
The default is “Normal”. Click the drop down box to switch the pulse polarity type
Normal: Perform pulse modulation when the pulse modulation signal is high.
Inverse: Do not perform pulse modulation when the pulse modulation signal is
high.
Pulse Mode
When the modulation source is “Int”, the SSG5000X provides three pulse modes:
“Single”, “Double” and “Train”. The default is “Single”. Click the drop
down box to select the pulse mode.
Single: A pulse signal is generated during one pulse period. At this time, the
two settings “Double Pulse Delay” and “#2 Width” are hidden.