Data Sheet

SPECIFICATIONS
Specifications are valid under the following condition: The instrument is within the calibration period, has been
stored between 0 and 50 for at least 2 hours prior to use, and has been powered on and warmed up for at
least 40 minutes. The specifications include the measurement uncertainty, unless otherwise noted.
Specifications: All products are guaranteed to meet published specifications when operating temperatures from
5 to 45
,
unless otherwise noted.
Typical (typ.): Performance deemed typical implies that 80 percent of the measurement results will meet the
typical published performance with a 95th percentile
confidence level at room temperature (approximately
20 ℃). Typical performance is not warranted and does not include measurement uncertainty.
Nominal (nom.): This value indicates the expected mean or average performance, or an attribute whose
performance is by design, such as the 50 Ohm connector.
Frequency characteristics
Frequency
Frequency range
SSG5040X
CW MODE 9 kHz-4 GHz
SSG5060X
CW MODE 9 kHz-6 GHz
SSG5040X-V
CW MODE 9 kHz-4 GHz IQ MODE 10 MHz-4 GHz
SSG5060X-V
CW MODE 9 kHz-6 GHz IQ MODE 10 MHz-6 GHz
Frequency resolution
0.001 Hz
Setting time
<5 ms (typ.), ALC ON
<10 ms (typ.), ALC OFF (S&H)
Resolution of phase offset
setting
0.1°
Frequency Band
[1]
Band
Frequency range
N
1
9 kHz f 1 MHz
0.25
2
1 MHz f 250 MHz
0.5
3
250 MHz f 500 MHz
0.125
4
500 MHz f 1000 MHz
0.25
5
1000 MHz f 2000 MHz
0.5
6
2000 MHz f 4000 MHz
1
7
4000 MHz f 6000 MHz
2
[1] N is a factor used to help define certain specifications within the document