SSG5000X User Manual 95
Target Level
Specify the nominal level expected at the input of the sensor. The signal generator
adjusts the output power accordingly, in order to meet the target value at the
sensor input, and thus the power required at the DUT.
The value can be set by the touch screen keyboard or the front panel keypad.
Level Limit
Set an upper limit for the RF output power to protect your DUT from damage due to
high input power. If the input RF signal power exceeds the limit, the set value will
not take effect and the source will pop up a warning message.
Capture Range
The capture range means that if the power sensor reading is within the effective
capture range, it is a valid reading and the RF signal is adjusted and compensated.
Set the capture range of the control system. If the power sensor reading exceeds
this range, the reading will be ignored.
Effective Capture Range = Target Level +/- Catch Range