2. Touch operation
Click on SHORTCUT->LEVEL->Level Offset to enter the interface of level offset
setting.
⚫ Click the level parameter, then pop up the keypad interface, then click the
number symbol in the keypad to change the value. Click the check mark to
confirm the modified value.
⚫ Long press edit box to enter the edit mode. Press the up and down direction
keys, rotate the knob or press the numeric keypad to modify the value, and
press the ENTER key, knob or ESC key to exit the edit mode.
Note:
⚫ The difference value between the display level and the level offset is the actual
RF output level, which must not exceed the instrument level output range.
ALC State
Setting the ALC’s working state
ALC is Automatic Level Control .It is an adaptive control system to stabilize the RF
output level.
It continuously monitors the current level and adjusts it to keep a steady state over