Laser Beam Analyzer OPERATOR'S MANUAL LBA-700/708/710/712/714PC

1 Standard Deviation
2 Minimum
3 Maximum
All results values will be loaded into the array all the time. Results not enabled in the LBA-PC will
be zero. Note there are three sets of Gauss Fit and Top Hat results. Either the whole beam
section or the major and minor sections will contain valid results depending on how the LBA-PC is
configured.
9.3.1.23 Pass/Fail Results
The PassFailFlag and PassFail properties provide all of the LBA-PC pass/fail test results.
The PassFailFlag is a combination of all the individual pass/fail results. The value of PassFailFlag
is defined as follows:
Fail Some test failed
No test No individual test enabled
Pass All tests passed
The PassFail property provides individual pass/fail results in a one-dimensional array of integers
packaged as a Variant. The pass/fail results are loaded into the array in the following order:
1. Quantitative
There is no test for Peak Location X or Y. The value of both of these flags is always zero.
There is no test for Radius so the flag is always zero.
2. Elliptical
3. Gauss Fit – whole beam
4. Gauss Fit – major axis
5. Gauss Fit – minor axis
6. Top Hat – whole beam
7. Top Hat – major axis
8. Top Hat – minor axis
9. Divergence
This is the same order as listed in sections 3.1.15 through 3.1.19 above under individual results
properties.
Each ‘flag’ will contain one of the following values:
-1 Fail
Operator’s Manual LBA-PC
168
0 No Test