Laser Beam Analyzer OPERATOR'S MANUAL LBA-700/708/710/712/714PC

2.6.1.2 Setup requirements for LBA-PC with pyrocam cameras:
Two files are provided for configuring the LBA to a Pyrocam I. They are ~PYROCAM.CFG and
~PYROCAM.CAM.
2.6.1.2.1 Setting up the Pyrocam Configuration.
Go to File. . . Restore Config. . . and set the configuration to ~PYROCAM.CFG.
2.6.1.2.2 Setting the camera type to Pyrocam
Go to Options. . . Camera dialog box and set the Camera selection to ~PYROCAM.CAM.
This may already have happened when you did the previous step.
The ~PYROCAM.CAM file is absolutely required for correct operation with your Pyrocam. The
~PYROCAM.CFG configuration is a good starting point configuration.
You will undoubtedly need to make changes to your configuration to suit your application.
Do so, and then save off your new configurations into new <filename>.cfg config files.
Remember the ~PYROCAM files are read only; so don’t try to use those file names.
2.6.1.3 Some Restrictions apply when interfaced to a Pyrocam I
2.6.1.3.1 Frame resolution restrictions
While you are not restricted from setting your frame Resolution to 512x480x1, it is not a
good idea. Keep your Resolution setting set to 128x120x4. Remember the Pyrocam creates
a 124x124 sized image, so setting higher resolutions only wastes memory and slows down
operations.
2.6.1.3.2 Zooming restrictions
In keeping with the above, do not use Hardware zooming. Only use Soft zooming, i.e.
double-right-mouse click in the Zoom window.
2.6.1.3.3 Restricted pixel regions
Because of the image size difference between the LBA and the Pyrocam I, as noted above,
the LBA will clip off 4 rows from the Pyrocam; two on top, and two on the bottom. Also the
LBA will have 4 dark columns; two on the left edge, and two along the right.
2.6.1.3.4 Scale and resolution restrictions
Note also that the Pixel Scale setting is set to 25mm, not 100mm, which is the actual pixel
scale of the Pyrocam I. This is because the image output by the Pyrocam in CCIR mode is 4
times over scanned in both the x and y directions. Thus the LBA’s 1x resolution comes out to
be 1/4th the 4x resolution. Keep this in mind if you plan to change the scaling because of
placing the beam altering optics in the beam path.
Operator’s Manual LBA-PC
28