Users Manual Part 2

Operation TP-R-400
32
2.5 Force setting (Measuring Unit PP41.00 only)
INFORMATION
In the case of strong vibrations in the machine or when using heavy probe inserts, an increase in the trigger force
may improve the measuring results.
1. Unscrew the stylus (refer to chapter 2.2.1).
2. Use an AF1.3 mm hexagon key to increase or decrease the trigger force (refer to Fig. 18).
Adjusting trigger force
3. Screw in the stylus (refer to chapter 2.2.1) and align the stylus to the spindle centre (refer to chapter 2.2.5).
4. Calibrate the probe.
Revolutions based on lowest
force setting
≈ Trigger force with 50 mm stylus in N
Z X/Y
min. 0
2.5
0.3
1
3.5
0.4
2
4.5
0.5
3
5.5
0.6
4
6.5
0.7
5
7.5
0.9
6
8.5
1
7
9.5
1.1
8
10.5
1.2
9
11.5
1.3
max. 10
12.5
1.4
Factory setting
Hexagon key AF 1.3 mm
(0227)