Users Manual Part 2

TP-R-400 Operation
33
2.6 Maintenance and Cleaning (Measuring Unit PP41.00 only)
NOTICE
Risk of damage to the equipment!
Clean and dry the probe well before opening!
Do NOT clean with compressed air or high pressure water!
Do NOT use any sharp tools (these could damage the inner seal)!
1. Unscrew the stylus.
INFORMATION
Dirt can collect under the metal eyelid.
2. Remove the service cover with metal eyelid and conical spring by hand.
3. Clean the touch probe and components under running water.
4. Reassemble all parts.
5. Align the stylus to spindle centre.
6. Calibrate the probe!
Maintenance and cleaning of the radio-wave touch probe TP-R-400
Service Cover (3240)
Metal eyelid
(2906)
Conical spring
(2931)
Inner seal
O-Ring 16x1 Viton (3455)