Specifications
ISTCD DPM, continued
Step Procedure
5. Remove DPM, along with its mounting bracket, using an
8 mm socket wrench or nut driver. There are two nuts
securing the metal mounting bracket to the enclosure. One
is visible in the front and one is at the back of the module.
6. If required, the DPM board can be removed from the bracket
by removing the two screws that mount the board to the
bracket (not the screws that connect the black cover
together). If required, the new DPM can be mounted on the
same bracket.
7. Verify that all external switch settings on the new board
match the old board exactly. Refer to chapter 2 of this
manual for further information relating to switch settings on
the ISTCD DPM.
8. Install the new board into the analyzer. Secure both nuts
using an 8 mm socket wrench or nut driver.
9. Connect the ground wire, taking care to reinstall the standoff
correctly as shown below.
10. Reconnect the I
2
C and Oven cables.
11. Close the door and reapply power to the device.
A5E31405710001 4-27