Specifications

Maxum II Specifications, Continued
Installation
Configuration
Single unit with multiple enclosures
Dimensions Height39 3/4" (1010 mm) [ Modular Oven - 24 11/16 (627 mm) ]
Width:26 1/16" (662 mm)
Depth: 16 3/16" (411 mm)
When mounting the analyzer on a wall, care should be taken to ensure that the
wall (vertical mounting surface) can withstand four times the minimum weight
of the analyzer when mounted with the appropriate hardware. In some cases,
it is recommended that brackets, such as Unistrut or angle iron be added to the
mounting surface to help distribute the weight.
CAUTION
Mounting Wall mount: center to center 44" (1120 mm)
Left side clearance:18" (460 mm)
Front side clearance:25 3/4" (654 mm)
Right side clearance: 18" (460 mm)
Weight
Airbath/Airless Oven Configuration
:
170 lb (77 kg) – typical, dependent on application
Modular Oven Configuration
:
150 lb (68 kg) – Maximum, may be less depending on application
Electronics Enclosure
Rating
Airbath/Airless: NEMA 3, IP44 Category 2
Modular
: NEMA 3, IP54 Category 1
EMI/RFI Rating CE Compliance; certified to 89/336/EC and 2004/108/EC (EMC directive)
CE Compliance; certified to 73/23/EC and 2006/95/EC (Low Voltage
directive)
Tested per EN 61010-1 / IEC 61010-1
Hazardous Class
Airbath/Airless Oven
Standard Configurations:
Certified by CSA C/US for use in Class I, Division 1, Groups B,C,D with air
or nitrogen purge
Certified by CSA C/US for use in Class I, Division 2, Groups B,C,D
Certified according to ATEX with air or nitrogen purge and purge control
for Zone 1 or Zone 2 (Ex pyedmib IIB + H2)
Suitable for use in general purpose and non-hazardous areas.
Important:
General Purpose, Division 2 and Zone 2 applications require
environmental purge of Electronic Enclosure (EC) to maintain operation
integrity and performance
PDHID is not rated for hazardous areas
1-8 A5E31405710001