Specifications
About The Maxum II, Continued
Parallel Chromatography can reduce the cycle time for complex
applications and also increase chromatograph analysis frequency by
running duplicate modules in parallel at staggered times. Since times are
staggered the system will provide more frequent measurement updates.
If similar measurements are performed on different streams, parallel
modules can be used for each stream instead of switching the stream to
a single module. This will reduce overall cycle time on multiple stream
applications.
Duplicate Modules
Use of parallel chromatography can reduce calibration requirements by
running two identical modules in parallel on the same stream to obtain
redundant measurements. As long as the results remain the same within
a predefined error limit, the analysis is known to be accurate. Deviations
outside the error limit can trigger notification or activate analyzer
calibration. Overall, the Maxum II calibration requirements are
significantly lower because of the parallel measurement configurations
and standard modular applications.
Redundant Measurements
Example
S S S R
Column 1
Column 2
1
2
10
9
8
3
4
5
7
6
Carrier In
from EPC
Sample
Out
Sample
In
SSO
Detector Vents
Fixed Restrictor
Figure 1-3: Applications Module Example
A5E31405710001 1-17