Specifications

Electronic Pressure Control (EPC) Module, Continued
As shown in the figure below, the EPC has a set of four DIP switches.
These are used to set the unique location ID for the specific EPC, which
is used in software as part of the hardware ID string. The location ID is
set using a binary counting of the switches from right to left (as
numbered on the board and not on the actual switches).
Location ID Switch
Settings
Location IDs for installed EPC boards count from the bottom up, with
ID#1 on the bottom and ID#4 on the top.
Figure 2-21: DIP Switches on the EPC
Location ID #1 Location ID #2 Location ID #3 Location ID #4
1
st
switch set
(binary 1)
Second switch
(binary 2)
1
st
& 2
nd
switch
set (binary 3)
3
rd
switch set
(binary 4)
Table 2-7: DIP Switches on the EPC
A5E31405710001 2-31