Specifications
General Maintenance, Continued
To gain access to the modules, the electronic enclosure door must be
opened. It will be necessary to use a #4mm Allen wrench to open the
door if the Allen screw on the latch has been tightened.
Opening Doors
After performing any maintenance function(s), check to be certain there
is no loose hardware left within the electronic enclosure. Such items can
create electrical shorts causing damage to internal components. This
increases system downtime for performing of corrective maintenance.
Inspection After
Maintenance
Recommended tools for performing maintenance are as follows:
Field Tool Kit
• Maxum II Tool Kit
or
• Set of metric Allen wrenches
• Set of metric wrenches or nut drivers
Note: Special tools required for specific procedures within this section
are noted within the procedure (example: torque wrenches
required for valve assembly).
It is important for proper procedures to be used when cleaning valve and
detector parts. All foreign contamination adhering to the part should be
removed using an appropriate cleaning solvent, such as hexane,
acetone, or methanol and a dust/lint free cloth. Use of an ultrasonic
cleaner is often helpful. After cleaning, it is necessary remove excess
cleaning fluid from the components by blowing with clean air or shaking.
Components must be air dry before reassembling.
Use of Solvents and
Detergents
It is also possible and often better to use an appropriate detergent, such
as Alconox®, for cleaning instead of solvent. However, after cleaning
with a detergent, it is necessary to rinse the part thoroughly with
deionized water (distilled water is also acceptable) in order to remove
detergent residue. All water must then be removed by blowing with
clean air or shaking. Components must be completely dry before
reassembling.
A5E31405710001 4-5