Specifications

Control Interface Module, Continued
The battery should only be replaced with an approved spare. Contact
Siemens for a replacement.
IMPORTANT
Battery Replacement
Step Procedure
1. Power off the Maxum and then open electronic enclosure
door, using a 4mm (5/32”) Allen wrench if necessary.
Before removal of battery, note location of its positive end when installed
in battery holder. The positive and negative terminals are marked on the
battery.
CAUTION
2. Remove the battery from its mounting bracket located on the
left side of the module. Refer to picture below.
3. When installing the new Lithium Battery in its holder, place
the positive (+) end so it installs per marking on the board
(positive terminal facing up).
4. Before closing door and reapplying AC power, be certain the
battery is securely installed in its holder and polarity, within
holder, is correct.
5. After battery is replaced, current date and time information is
lost. If the analyzer is configured to obtain date and time
information from a central server, then it will update
automatically. If no time server is set, it will be necessary to
manually set the date and time on the analyzer.
Batter
y
Figure 4-1: CIM Board Battery Location
4-12 A5E31405710001