Instruction Manual Model ISC402G Inductive Conductivity Converter IM 12D8C2-E-E 6th Edition
OUTPUT HOLD FLAG MEASURED VALUE DISPLAY FAIL FLAG HOLD FAIL MENU POINTER FLAGS MENU FOR MAINTENANCE FUNCTIONS SEE CHAPTER 5 MODE ∝S / c m mS/cm MESSAGE DISPLAY YES NO ENT MEASURE SETPOINTS CALIBRATE DISPLAY HOLD AIR SET SETPOINTS RANGE SET HOLD TEMP.
Code Function Use Default 01 02 03 04 05 Temperature sensors and units Installation factor Weight indication Output range Output function 0.0 1.88 0.0 1.
Chapter Description INSTRUMENT DESCRIPTION AND CONTROLS Page Front cover SERVICE SETTINGS Front matter CONTENTS Front matter 1. INTRODUCTION 1-1. Application..........................................................................................1 1-2. Measurement principle........................................................................1 1-3. Functional description ........................................................................2 1-4. Instrument check ...............................
Chapter Description 7. Page CONFIGURATION IN SERVICE MODE Introduction ..............................................................................41 7-0. Access to service settings ........................................................42 7-1. Temperature sensor & units ......................................................43 7-2. Installation factor adjustment ....................................................43 7-3. Weight % on second display line ..............................................
1. INTRODUCTION 1-1. Application The model ISC402G inductive conductivity mains powered converter is designed for use with the Model ISC40 sensor. This revolutionary conductivity measuring system features ± 0.5% accuracy over a wide range of conductivity values (1 to 2,000,000 µS/cm) and process temperatures (-30 to 150 °C/22 to 302 °F) without changing cell constant or recalibration.
2 specific conductivity and a constant factor that is determined by the geometry of the sensor (length divided by surface area of the hole in the toroid) and the installation of the sensor. There are 2 toroids mounted in the "doughnut" shaped sensor. The liquid also flows through the second toroid and therefore the liquid turn can be considered as a primary winding of the second ring transformer. The current in the liquid will create a magnetic field in the second toroid.
3 2. SPECIFICATIONS 2-1. General specifications A. Input specifications : Yokogawa Model ISC40 inductive conductivity sensor with integrated temperature sensor YSI thermistor PT 1000 Ω. B. Measuring range Conductivity E. Transmission range Conductivity : Minimum conductivity at process temperature : Maximum conductivity at process temperature : Temperature : C. Indicating range Main display Message display D.
4 Conductivity alarm : Temperature alarm : Adjustable hysteresis and delay time Proportional duty cycle control with adjustable proportional range, pulse period and fixed minimum/maximum duty cycle. Proportional frequency control with adjustable proportional range and pulse frequency. Adjustable hysteresis and delay time. Fail contact : Signals a fail of the measuring loop. : For storing measuring/setup data. G. Logbook H. Power supply Model ISC402G I.
5 2-3. Environment and operational conditions: A. Ambient operating temperature : -10 to + 55 °C (10 to 131 °F) Excursions to -30 °C (-20 °F) do not influence the current output function and excursions to +70 °C (160 °F) are acceptable too. C. Relative humidity : 10 to 90 %. D. Weather protection : Rain and dust tight to IP 65 (NEMA 4X). E. Data protection : F. Watchdog timer : G. Automatic safeguard : H.
6 E. Mounting configurations : Universal mounting kit available as an option for wall and pipe mounting. Separate mounting kit available as an option for panel mounting. Suitable for DIN sized panels 144 x 144 mm (5.67 x 5.67 inch). F. Shipping details Dimensions (wxhxd) : 144 x 144 x 132 mm (5.7 x 5.7 x 5.2 inch). 290 x 225 x 170 mm (11.5 x 8.9 x 6.7 inch). Approximately 2.5 kg (5 lb). 2-6. Model and suffix codes Package (wxhxd) : Weight : IM 12D8C2-E-E 2-5.
7 3. INSTALLATION AND WIRING 3-1. Installation and dimensions The EXA ISC402G should only be used in conjunction with apparatus which meets the relevant IEC, American or Canadian standards. Yokogawa can accept no responsibility for the misuse of this equipment. 3-1-2. Mounting methods The EXA ISC402G transmitter has universal mounting possibilities: - Panel mounting using optional brackets - Surface mounting on a plate (by bolts from the back) - Wall mounting on a bracket (e.g.
8 Wall Mounting Pipe Mounting (Vertical) Pipe Mounting (Horizontal) 56 (2.20) 200 (7.87) 2x ø6.5 (0.26) 4x ø10 (0.4) 70 (2.75) 145 (5.70) 2" I.D. pipe OPTION/U: Universal pipe/wall mounting kit Figure 3-3.
9 For easy wiring the sensor must be located within 5/10 m (16/32 ft) from the transmitter using the integral sensor cabling (refer to fig. 3-4). Unit: mm (inch) Dimensions L=20000 (800) L= 5000 ( 200 ) Installation instructions bulk-head mounting 11 12 17 13 15 16 14 t wrench opening 20 ( 0.79 ) 58 (2.28 ) wrench opening 32 ( 1.42 ) A Ø 23.5 ( 0.92 ) D Ø 27 ( 1.06 ) I Ø 40 ( 1.57) 124 (4.88 ) 3-1-3. Installation of the sensor The ISC40G is a doughnut shaped sensor.
10 3-2. System configuration FRONT GLANDS REAR GLANDS SENSORS 0/4-20mA OUTPUT SIGNALS POWER CONTACT OUTPUT 0/4-20mA S1 S2 RS485 CONTACT OUTPUT S3/WASH S4/FAIL CONTACT INPUT Figure 3-5.
11 3-3. Wiring 3-3-1. Preparation The relay contact terminals and power supply connections are under the screening plate. These should be connected first. Connect the input, output and databus connections last. The procedure to open the EXA ISC402G is described below: 1. Loosen the four captive screws and remove the cover. 2. Use the rubber knob and swing open the display board to the left. 3. The upper terminals strip is now visible. 4. Remove the screen plate covering the lower terminal strip. 5.
12 High voltage section Sensor Contact (S3, S4,Fail) cable(s) output cable(s) Contact (S1, S2) output cable(s) Analog output cable(s) Power cable Communication,Con -tact Input Figure 3-7. Glands for cabling Option tag Figure 3-8. Grounding the housing IM 12D8C2-E-E Figure 3-9.
13 3-3-2. Wiring of power supply 3-3-2-1. General precautions Make sure the power supply is switched off. Make sure that the power supply is correct for the specifications given. Remove the front cover by unscrewing the 4 captive screws. Check the type-plate on the instrument for the correct supply voltage. Local health and safety regulations may require an external circuit breaker to be installed. The instrument is protected internally by a fuse.
14 3-3-4. Wiring the contact signals 3-3-4-1. General precautions The contact output signals consists of voltage-free relay contacts for switching electrical appliances (SPDT). They can also be used as digital outputs to signal processing equipment (e.g. PLC). It is possible to use multi-core cables for the contact in-and output signals and screened multi-core cable for the analog signals. 3-3-4-2. Contact outputs The 4 contact outputs can be wired to your own requirements, refer to Fig. 3-7.
15 4. START-UP The EXA ISC402G transmitter offers great flexibility to let the user have one hardware version for a multitude of different applications. The software has been programmed for a "general application". It will be necessary for the user to program the instrument to his specific requirements or application. Although programming the transmitter is simple, some preparation is necessary.
16 4-3. Temperature (TEMP) compensation The factory adjustment for automatic temperature compensation is the standardized NaCl algorithm according IEC 746-3. For concentrated electrolytes this may not be accurately compensate for process temperature variations. The ISC402G gives the user 4 methods of optimizing the temperature compensation: 4-3-1.Field calibration of (linear) temperature compensation The procedure is described under §6-4. 4-3-2.
17 4-6.
18 4-7. Cleaning If it is necessary to clean the outside of the EXA housing, this should done using only a soft cloth and household detergent. DO NOT USE corrosive or abrasive cleaners. 4-8. Battery The EXA instrument contains a lithium cell to support the clock function when the power is switched off. This cell needs to be replaced at 5 yearly intervals (or when discharged). Contact your nearest Yokogawa service centre for spare parts and instructions.
19 5. CONFIGURATION IN MAINTENANCE MODE 1. ACCESS TO CALIBRATION ROUTINE 5-1. CALIBRATION 2. ADJUST VALUE MODE MODE µS/cm YES NO MEASURE SETPOINTS CALIBRATE DISPLAY HOLD NO MODE AIR SET SETPOINTS RANGE SET HOLD TEMP.
20 1. When is calibration necessary? Initially follow the procedure given in item §4-4-1 for air calibration when setting up the instrument for the first time. Calibration of conductivity instruments is not normally necessary as the conductivity cells are manufactured to close tolerances and do not alter in use. If the cell has severe fouling or been subject to abrasion (possibly during cleaning) it may be necessary to calibrate.
21 With this method the sensor is not removed from the process. This method is the most convenient method of calibrating the ISC402G transmitter. Since the sensor is immersed in the process, errors caused by installation characteristics are compensated for. NOTE: The standard instrument used as reference method must be accurate. Yokogawa recommends that the Model SC82 pocket conductivity meter be used for this purpose. 3.
22 5-2. SELECTING A VALUE TO DISPLAY 2. SELECT DISPLAY 1 PARAMETERS 1. ACCESS DISPLAY 1 OR DISPLAY 2 MODE YES NO MODE MEASURE SETPOINTS CALIBRATE DISPLAY HOLD NO MODE AIR SET SETPOINTS RANGE SET HOLD TEMP. SERVICE CONTACTS S1 * YES YES NO MEASURE SETPOINTS CALIBRATE DISPLAY HOLD NO MODE > MODE Press MODE key for access to maintenance mode NO YES Continue with step 2 for DISP.1 parameters. Continue with step 3 for DISP.2 parameters.
23 5-2. SELECTING A VALUE TO DISPLAY 1. What is a display routine? The second line of the display is to show: - Actual status - Messages - Errors. When delivered from the factory the EXA ISC402G shows the temperature on the second line. You can make the instrument show a different parameter on the second line by selecting it from the list on the right. NOTE: DISP.2 is only activated when % concentration or Temp. comp. type are set differently for each mA output. 2.
23 5-3. USE OF HOLD FUNCTION 1. ACCESS TO HOLD ROUTINE 2. SWITCH HOLD ON/OFF HOLD MODE YES YES MEASURE SETPOINTS CALIBRATE DISPLAY HOLD NO MODE NO AIR SET SETPOINTS RANGE SET HOLD TEMP.
25 5-3. USE OF HOLD FUNCTION 1. What is HOLD? Hold is a function which freezes the output signal temporarily, it is normally used during maintenance to prevent unwanted controller reaction when the cell is removed from the solution being measured. The HOLD function must be commissioned from the programming menu before it can be switched on or off. See commissioning the hold function for more details. NOTE: AUTO RETURN will switch off HOLD after 10 minutes, see §7-10. 2.
26 5-4. ADJUSTING THE SET POINTS 1. ACCESS TO SET POINTS MENU 2. SELECT SET POINT TO ADJUST MODE 3. ADJUST THE SET POINT VALUE MODE MODE mS/cm MODE YES MEASURE SETPOINTS CALIBRATE DISPLAY HOLD NO CONTACTS S1 * YES NO MODE > > S3 > ENT ENT FAIL/S4 NOTE: This function can only be used if activated during commissioning (See §6-3 and §7-18). MODE Press MODE key for access to maintenance mode.
27 5-4. ADJUSTING THE SET POINTS NOTE: It is only possible to access this function at the maintenance level if it has been activated with service code 24. This function gives the operator the facility to change the setpoints at the maintenance level. It can only be used if it has been activated in service code 24. The protection of the set points is to prevent unauthorized access to these functions. NOTE: 1.
28 6. CONFIGURATION IN COMMISSIONING MODE 1. ACCESS TO SET POINTS MENU 2. SELECT SET POINT TO ADJUST 3.
29 6-1. ADJUSTING THE SET POINTS 1. What are setpoints? In general the contact outputs of the EXA ISC402G can be used to signal extraordinary situations in the process. The alarm contacts are switched if the comparison of the changes process value to a fixed setpoint meets the condition of the alarm function (higher or lower). These setpoints are programmed in this section. One of the advantages of a micro-processor instrument is, that it is fairly simple to change the function of the contact outputs.
30 6-2. OUTPUT RANGE ADJUSTMENT 2. ADJUST START VALUE OF THE RANGE 1. ACCESS TO OUTPUT MENU 3. ADJUST THE END VALUE OF THE RANGE MODE MODE MODE µS/cm mS/cm NO MODE ENT MEASURE SETPOINTS CALIBRATE DISPLAY HOLD CONTACTS S1 * NO YES MODE CONTACTS S1 S2 FAIL/S4 Remove cover by releasing 4 screws ⊗ Press this key for access to commissioning mode Select RANGE menu NO S3 > > > ENT YES NO YES MODE Press YES-key to adjust RANG.1 or press NO-key to change to RANG.
31 6-2. OUTPUT RANGE ADJUSTMENT 1. What is the output range? The default value of the software defines the output range of 4-20 mA = 0-1000 mS/cm. This display will always show the full range of the instrument from 0-2000 mS/cm. Maximum resolution is achieved by auto-ranging, where the position of the decimal point and measuring units (microSiemens/milliSiemens) are moved to best fit the actual value. For control or recording the current output can be ranged to a specific part of the total measuring span.
32 6-3. SET UP HOLD FUNCTION 2. SELECT SET POINT TO ADJUST MODE YES MEASURE SETPOINTS CALIBRATE DISPLAY HOLD NO NO MODE AIR SET SETPOINTS RANGE SET HOLD TEMP. SERVICE CONTACTS S1 * YES YES NO MEASURE SETPOINTS CALIBRATE DISPLAY HOLD NO MODE S1 > S3 > ENT FAIL/S4 FAIL/S4 Remove cover by releasing 4 screws Display shows actual status *H.OFF = Hold not activated *H.
33 6-3. SET-UP HOLD FUNCTION 1. What is HOLD? HOLD is a function freezing the output signal temporary, during normal maintenance, preventing unintended alarm or colour actions to occur. Two possibilities are generally used: - Keep the output at the LAST (H.LST) value just before the start of maintenance. This can only be used when a recorder is connected. - Keep the output at preset FIXED (H.FIX) value which will not cause any of the alarms to go off or any controlling action to be taken.
34 6-4. TEMPERATURE COMPENSATION 2. SELECT STANDARD OR MANUALLY SET COMPENSATION 1. ACCESS TEMPERATURE COMPENSATION ROUTINE MODE 3. SET THE TEMPERATURE COMPENSATION COEFFICIENT (only if T.C.
35 6-4. TEMPERATURE COMPENSATION 1. Why is temperature compensation necessary? The conductivity of a solution is very strongly influenced by temperature. Typically for every 1 °C change in temperature, the solution conductivity will change by approximately 2%. The effect of temperature varies from one solution to another and is determined by several factors; solution composition, concentration and temperature range.
36 4. Calculating the approximate temperature coefficient If the temperature coefficient of a liquid to be measured cannot be found, measure the conductivity of the liquid at two temperatures between 10 and 30°C with the temperature coefficient set to 0.00 and use the following equation to calculate an approximate temperature coefficient (α). K2 - K1 Temp.coef.
37 IM 12D8C2-E-E
38 6-5. AIR SET 2. ACTIVATE AIR SET FUNCTION MODE YES YES NO MODE AIR SET SETPOINTS RANGE SET HOLD TEMP.
39 6-5. AIR SET 1. What is AIR SET procedure? The AIR SET procedure is a routine to compensate for the small zero current due to the inductive coupling of the sensor, cabling and transmitter. 2. When is an AIR SET procedure necessary? AIR SET is done only once during Start-Up, and when a new sensor is connected to a transmitter. If the user wants to make full use of the systems capabilities and requires accurate measurement down to 0.5 µS/cm, this simple routine will optimize the performance. 3.
40 IM 12D8C2-E-E
41 7. CONFIGURATION IN SERVICE MODE Introduction Usually there is no requirement to adjust the settings of the service section. All parameters are pre-programmed to default values which lets the instrument be put into service immediately. The advanced functions available through this section are only needed in some specific applications. This fine-tuning of the instrument gives a superior performance over conventional instruments.
42 7-0. ACCESS TO SERVICE SETTING 1. ACCESS TO SERVICE MENU 2. ENTER CODE TO SELECT THE REQUIRED SERVICE FUNCTION MODE MODE AIR SET SETPOINTS RANGE SET HOLD TEMP. SERVICE CONTACTS S1 * ENT NO YES MODE > ENT Press this key for access to commissioning mode NO Select calibration routine Press NO-key until display indicates SERV.
43 Temperature in °C Temperature in °F NTC PT1000 Explanation: The code for the temperature units determines the temperature indication on the display. The sensor selection chooses between the integral (NTC) sensor, and external PT 1000 Ω RTD. Default: NTC. °C. * put PT 1000 Ω RTD on one line Adjustment: Adjust the ratio between the measured conductance of the sensor and the specific conductivity of the solution.
44 7-3. Weight % on second display line ACCESS CODE : 03 (see §7-0 for operation) NOTE: This routine is only useful if the output range in code 05 is linear to weight %. DISPLAY : *% Adjustment : X.X X.X 0.X 1.X X.0 X.1 OFF ON OFF ON %.1 %.1 %.2 %.2 Select the possibility to display % by weight on the second display line.
45 7-5. Output function ACCESS CODE : 05 (see §7-0 for operation) DISPLAY : OUTP. F Adjustment : X.X 0.X 1.X X.0 X.1 X.2 X.X Linear range for mA 1. Output table for mA 1. Linear range for mA 2. Output table for mA 2. Temperature output for mA 2. Explanation: As a default, a linear output is set and only 0% and 100% can be adjusted from the commissioning output function. When an output table is programmed by setting a 1, the table can be programmed through code 06.
46 % Output OUTPUT LINEAR TO CONCENTRATION EXAMPLE 0-25% SULFURIC ACID Conductivity(mS/cm) 1000 Output in % 100 800 80 600 60 400 40 200 20 0 0 0 2 4 6 8 10 12 14 16 18 20 22 24 Concentration(% by wt) Figure 7-2. IM 12D8C2-E-E Conductivity % Output mA Concentration Example Conductivity Example 0 4 0 0 5 4.8 1.25 60 10 5.6 2.5 113 15 6.4 3.75 180 20 7.2 5 218 25 8 6.25 290 30 8.8 7.5 335 35 9.6 8.75 383 40 10.4 10 424 45 11.2 11.25 466 50 12 12.
47 7-7. Contacts S1 to S4 ACCESS CODE : 08, 09, 10 and 11 (see §7-0 for operation) DISPLAY : *S1, *S2, *S3 and *S4 For *S1, *S2 and *S3 Adjustments : X.X 0.X 1.X 2.X 4.X X.0 X.1 X.2 X.3 X.
48 7-8. Settings for proportional control ACCESS CODE : 12, 13 and 14 (see §7-0 for operation) DISPLAY : *RANGE, *PER., *FREQ. Adjustment: The proportional range (*RANGE) is set in % of the programmed output range. The proportional pulse period (*PER) is set in seconds. The proportional pulse frequency (*FREQ) is set in pulses per minute. Explanation: It is only useful to set these values when one or more contacts will be used for proportional control.
49 7-9. Setting the process alarm function ACCESS CODE : 15 (see §7-0 for operation) DISPLAY : *D.TIME and *HYST Adjustment: Setting the delay time (in seconds) and hysteresis (in percent of setpoint). Explanation: The delay time is counted from the moment that the process value has passed the setpoint of the alarm function, refer to Fig. 10-3.
50 7-11. Signalling of fail condition ACCESS CODE : 17 (see §7-0 for operation) DISPLAY : *BURN 7-12. Temperature adjustment ACCESS CODE : 18 (see §7-0 for operation) DISPLAY : *T.ADJ Adjustment : Setting the measured temperature to a calibrated value, using the >, ^ and Ent keys. Adjustment : (X) X 0 1 No special output on FAIL. Special output signals on FAIL activated.
51 7-13. Temperature coefficient ACCESS CODE : 19 (see §7-0 for operation) NOTE: Access to this routine is only possible if "TEMP 1" or "TEMP 2" routine is set to "T.C." in section 6-4. DISPLAY: *T.C.1 and/or T.C.2 Adjustment: Adjust the value of the compensation factor (α). Explanation: In addition to the procedure described in §64 it is possible to adjust the compensation factor of the sample liquid if the compensation factor is known from laboratory experiments or has been previously determined.
52 7-15. Timer on contact function ACCESS CODE : 21 (see §7-0 for operation) DISPLAY : *EXPIR/*tE.min 7-16. Setting the clock ACCESS CODE : 22 (see §7-0 for operation) DISPLAY : *HOUR Adjustment : Adjust the value for the reference temperature used in the calculations. Normally this value will be 25 °C. Adjustment : (X) Adjustment: Set the internal clock to the current time. Use the >, and ENT keys to program the current time. The sequence is Hours, Minutes, Seconds, Year, Month, Day.
53 7-17. Settings for the temperature hysteresis ACCESS CODE : 23 (see §7-0 for operation) DISPLAY : *T.HYST 7-18. Enable Setpoint adjustment in maintenance ACCESS CODE : 24 (see §7-0 for operation) DISPLAY : *MODE Adjustment: Settings the hysteresis for setpoint of contact allocated to process temperature. Adjustment: (X) Explanation: When dealing with a high [low] alarm the relay will switch back after the process value has passed the setpoint minus [plus] hysteresis.
54 7-20. Fail status signalling ACCESS CODE : 26 (see §7-0 for operation) DISPLAY : *Err.X *SOFT 7-21. Communications setup ACCESS CODE : 27 (see §7-0 for operation) DISPLAY : *COMM Adjustment: Adjustment: (X) (for each error code number displayed). X 0 1 Error triggers flashing FAIL flag and LED and pulsating FAIL contact. Error triggers continuous FAIL flag and LED, continuous FAIL contact and, if activated, a 22 mA output signal. All other contacts will be deactivated.
55 7-22. Logbook clear ACCESS CODE : 28 DISPLAY : LOG.CRL Explanation: This function is only available in combination with the communications output package. This function is used to set the level of event logging, approximately 100 events can be stored in memory. When the memory is full, new events are logged by writing over the oldest entry. 7-23. Selection of standard electrolytes ACCESS CODE : 30 (see NOTE) (see §6-4 for operation) DISPLAY : *MATRIX Adjustment : Refer to table below.
56 7-24. Matrix Temperature compensation ACCESS CODE : 31, 32, 33, 34, 35 and 36 (see NOTE) (see §7-0 for operation) NOTE: These codes can only be accessed if temperature compensation in §6-4 is set to "MATRIX" and if code 30 is set to 9. In these service codes the ISC402 transmitter can be tailored for one particular application.
57 1. In defining the temperature compensation range, it is important to choose T1 and T5 in such a way that the 5 reference temperatures are user friendly: Example: T1 = 0 and T5 = 80 °C gives reference temperatures of 0, 20, 40, 60 and 80 °C. The minimum span for the range is 50 °C. 2. Use the table in the manual to record your programmed values. It will make programming easier for duplicate systems or in case of data loss. SODIUM HYDROXIDE 0 - 15% at 0 - 100°C 1200 1000 x mS/cm 800 NOTES: 1.
58 7-25. Built-in Test function ACCESS CODE : 50 (see §7-0 for operation) DISPLAY : *TEST, 7-26. Restore default settings ACCESS CODE : 55 (see §7-0 for operation) DISPLAY : *ERASE Explanation: This function tests the transmitter. Do the following procedure : Adjustment: Press ENT. Press ENT. Press ENT and repeat until display shows 22 mA. > When the display indicates 22 mA press ENT. Press the key. gled Use the > key to select * S2, S3 and S4 and repeat the test. After testing S4, Press ENT.
59 7-27. Data protection by three digit code ACCESS CODE : 66 (see §7-0 for operation) DISPLAY : *PASS Adjustment: X.X.X 0.-.- (X.X.X) Protection on Maintenance level not activated Protection on Commissioning level not activated Protection on Service level not activated -.0.-.-.0 #.-.- Protection on Maintenance level activated Protection on Commissioning level activated Protection on Service level activated -.#.-.-.
60 8. TROUBLE SHOOTING 8-1. Introduction The EXA ISC402G microprocessor based conductivity analyzer continuously monitors the condition of all key components of the measuring system to ensure that measurement is reliable. If a fault is detected it is immediately signalled. Errors are shown on the display with a code. The following table shows the errors which can be detected and gives information to help locate the fault or identify the error.
61 8-4. Error messages and explanation Code Error description Possible cause Remedy E2 Temperature coefficient incorrect. Calculated TC not between 0.00 and 3.50%/ºC. TC too high for measured temperature. See §7-13 E3 Installation Factor out of range.
62 Continuation from page 61. Code Error description Possible cause Remedy E20 Data integrity not good. Unauthorized programming Software problem The unit has to be re-initialized Call Yokogawa E21 Corrupted EPROM Software failure Call Yokogawa E22* Expired contact action interval Ineffective process control Check control loop * These errors will trigger the FAIL if set to on (default is on).
63 SPARE PARTS Description Part Number Pos. Number Cover assembly, including 4 pcs. M4 x 20 screws K1541JG 1 Digital and display board assembly * K1543DA 2 EEPROM with software (inc. label) K1543BG 3 Protective cover for power terminals (inc.
64 EXPLODED VIEW 11 2 12 10 5 1 8 13 6 4 3 Figure 8-1.
65 ERROR MESSAGES CODE E2 E3 E4 E5 E6 E7 E8 E9 E10 E15 E17 E18 E19 E20 E21 E22 DESCRIPTION TEMPERATURE COEFFICIENT INCORRECT INSTALLATION FACTOR OUT OF RANGE IMPOSSIBLE PROGRAM FOR TEMPERATURE COMPENSATION CONDUCTIVITY TOO HIGH CONDUCTIVITY TOO LOW TEMPERATURE TOO HIGH TEMPERATURE TOO LOW NO AIR SET POSSIBLE WRITING TO EEPROM FAILED TEMPERATURE CORR.
CODED SERVICE SETTINGS (DEFAULTS) CODE 01 02 03 04 05 DISPLAY T.CODE I.F. % mA TABLE 06 07 08 09 10 11 12 13 14 15 0% / 100% 0% / 100% S1 S2 S3 S4 RANGE PER FREQ D.TIME HYST RET BURN T. ADJ T.C. 1 / T.C. 2 T.R. °C / °F EXPIR HOUR/MINUTE/ SECOND/YEAR/ MONTH/DAY T.HYST MODE EN. E6 EN. E(5/6/7/8/22) *SOFT COMM LOG.CLR MATRIX T1....T5 °C / °F L1XT1......L1XT5 L2XT1......L2XT5 L3XT1......L3XT5 L4XT1......L4XT5 L5XT1......
IM 12D8C2-E-E EXA Series Model IC402 Converter for Inductive Conductivity Line Number ATE (automatic test equipment no.) Month code Year code Resistance decade boxes with ranges of 10 Ω - 100 kΩ and 0.1 Ω - 10 kΩ with an accuracy of 0.1%. Two 600 Ω resistors. An inductive conductivity sensor (IC40) complete with 5 metres of cable. 2 An insulated wire with a length of approx 1 metre and a minimum cross-section of 0.
IM 12D8C2-E-E Figure 1 (*AIR) appears in the lower display (*START) appears in the lower display (*WAIT) appears in the lower display, then (* END) appears (* AIR) Press YES. (*CODE) and 00 is dislayed Select code 50 and press ENTER (*TEST) appears Press YES (*CAL) appears Press YES and adjust the reading to 1.000 mS/cm This completes the system calibration.
IM 12D8C2-E-E Page 3 of 5 1.002 mS/cm ± 0.006 mS/cm ± 1.0 µS/cm ± 0.6 µS/cm + 0.5 µS/cm 1.000 mS/cm 100.0 µS/cm 10.0 µS/cm 0.0 µS/cm 1 kΩ 10 kΩ 100 kΩ (Open) ∞ Ω * -10 °C 25 °C 100 °C 158 kΩ 30 kΩ 2069 Ω ± 0.2 °C ± 0.2 °C ± 0.2 °C Tolerance 99.8 °C 24.9 °C -10.1 °C Actual Display * -10 °C 25 °C 100 °C 960.7 Ω 1097.3 Ω 1385 Ω ± 0.2 °C ± 0.2 °C ± 0.2 °C Tolerance 100.0 °C 25.1 °C -10.0 °C Actual Display * * Values in italics are shown for example only.
Page 4 of 5 ± 1 mS/cm ± 3 mS/cm ± 4 mS/cm ± 5 mS/cm 200 mS/cm 500 mS/cm 800 mS/cm 1000 mS/cm 500 Ω 200 Ω 125 Ω 100 Ω 1001 mS/cm 799 mS/cm 501 mS/cm 201 mS/cm 0.0 µS/cm 20.00 16.00 10.00 4.00 0.00 20.00 16.80 12.00 7.20 4.00 Nominal mA 0-20 or 4-20 Actual Reading * ± 0.13 ± 0.11 ± 0.08 ± 0.05 ± 0.03 mA Tolerance 20.00 16.79 12.01 7.21 4.00 mA1 * Actual 20.00 16.79 12.00 7.19 4.00 mA2 * Actual ± 0.02 ± 0.02 ± 0.02 ± 0.02 ± 0.02 8.0 12.0 16.0 20.0 22.0 22.01 20.
EXA Series Model IC402 Converter for Inductive Conductivity Line Number ATE (automatic test equipment no.) Month code Year code Resistance decade boxes with ranges of 10 Ω - 100 kΩ and 0.1 Ω - 10 kΩ with an accuracy of 0.1%. Two 600 Ω resistors. An inductive conductivity sensor (IC40) complete with 5 metres of cable. 2 An insulated wire with a length of approx 1 metre and a minimum cross-section of 0.
YOKOGAWA HEADQUARTERS 9-32, Nakacho 2-chome, Musashinoshi Tokyo 180 Japan Tel. (81)-422-52-5535 Fax (81)-422-55-1202 www.yokogawa.com YOKOGAWA CORPORATION OF AMERICA 2 Dart Road Newnan GA 30265 United States Tel. (1)-770-253-7000 Fax (1)-770-251-2088 www.yokogawa.com/us YOKOGAWA EUROPE B.V. Databankweg 20 3821 AL AMERSFOORT The Netherlands Tel. +31-33-4641 611 Fax +31-33-4641 610 www.yokogawa.com/eu YOKOGAWA ELECTRIC ASIA Pte. Ltd. 5 Bedok South Road Singapore 469270 Singapore Tel.