6.15 User Programmable Alarm System
The Synergy Nano’s user programmable alarm system can create customer specific alarms and warnings
and special factory applications. The user can create alarms for RTD temperatures, UUT temperatures as
well as voltage inputs, digital inputs and auxiliary sensors such as Oxygen, pressure, etc. For special
applications, the chamber manufacturer or retrofit installer can program an alarm relay to operate a system
function. This section explains the user alarm setup procedures using two examples.
The User Alarm screen operates as a wizard like the profile wizard screen. This means that the setup
instructions are provided along with the entry fields and the user enters information and presses next or back
as required until the alarm is entered. The result alarm can be programmed to appear in the Alarm window if
required as shown below. The diagram below right shows the electrical connection for the corresponding
user alarm.
There are 102 inputs/variables that can be monitored for user alarm conditions (see the table below)
The Input options are:
Module First Selection Second Sel. Choices
Olympic Board RTD1&2, Analog 1-4 N/A 6
UUT Module Inputs UUT Module Sensor 64
Machine Inputs Low Resolution Channels 1 thru 8 N/A 8
Digital Inputs Inputs 1 thru 16 N/A 16
Channels Channels 1 thru 4 N/A 4
Setpoints Setpoints 1 thru 4 N/A 4
There are four Comparison options available.
Comparison Application
Input Open Digital Inputs only
Input Closed Digital Inputs only
Greater than Threshold All inputs except Digital, Raw or Scaled.
Less than Threshold All inputs except Digital, Raw or Scaled.