Specifications
Synergy Nano Technical Manual, Revision -
Page 179
Loading, Editing and Saving Lists
To open a Settings List press the Load List button. Select a file and press OK. The name of the loaded file
is displayed on the title bar. The contents of the list appear in the Settings List data grid. The Settings
column holds the commands. The Value column holds the values for the commands. The Set Response
column holds the chamber’s responses from set commands.
To edit an item in the list, double click on the cell. The text is displayed in bold font when it is editable. Type
in your changes then hit Enter. If you don’t press Enter your changes will not be saved. If you need to add a
row to the end of the list, press the Add Row button. If you need to delete a row from the list, select the row
and press the Del Row button.
You can add to the Memo field by clicking in the Memo box and typing additional lines.
You can clear the entire list, including commands, by pressing the Clear List button. Alternatively you can
clear selected columns by pressing the Del Col button. When you are done editing the Settings List, save it
by pressing the Save List button.
Retrieving Data
To retrieve the current configuration of a chamber, load a Settings List that
contains the applicable set of commands. The commands do not require the
query “?” or set “=” operators. Once loaded, clear the Value and Set Response
columns if necessary. Make sure you are connected and press the Get button.
SimpleComm will query the chamber one command at a time until it has gone
through the entire list. After each query is sent to the chamber, SimpleComm
will wait until it receives a response before moving on to the next query. As
each response arrives it is placed in the corresponding cell in the Value column.
When the entire list has been transferred a popup window will state that the
transfer is complete.
Get Settings Example:
\\Serial Number: 06/0201
\\Version: 1.4.10
CF
CAL1
A1L