Specifications

Synergy Nano Technical Manual, Revision -
Page 215
Open File: Press this button to load a file from the Storage Card, USB Hard Disk or Floppy
Drive. The procedure is identical to the one described in Section 9.1 Programming the Synergy:
Loading a File. Simply select the desired file from the Drive List / File List screen that appears.
Run: Press the Run button to start the loaded program at the first step.
Run From: Use this button to start the program from a step other than the first step. Click on the
desired step to select it and press the Run From button.
Run Off: Use this button to run your program with the chamber outputs OFF. You can watch the
setpoint values change as the program runs to verify the program performs as expected.
Stop: Stops program execution.
Pause: Pauses program execution. The chamber will continue to run at the current setpoints.
Dyn. Edit: Dynamic Edit allows the user to edit the jump loop count while the program is running.
Dynamic Edit
To dynamically edit a Jump Loop the program must
be running and currently within the target loop.
Press the Pause button to pause the program.
Highlight the desired Jump Loop step then press
the Dyn. Edit button.
Press the Jump Loops Remaining text box to
open a numeric keypad and enter the new number
of jump loops to perform and press OK.
Press Run or Run Off to resume running the
program.
Note: Enter –1 to exit the loop.
Note: If the chamber loses power while running a program, the Synergy Nano is designed to resume the
program when power is restored. It will continue in the same step is was in when the power failed. If the
step was time based, such as a Waitfor or a Setpoint step, all timers are reset to zero and the Synergy Nano
resumes the profile at the beginning of the step.