UUT Module Setup Procedure
This is the installation and set up procedure for one or more UUT (Unit-Under-Test) modules. If a UUT
module is already installed in your chamber, go to Step 9 in this Procedure for instructions on how to view
your thermocouple temperature readings on the Synergy Nano touch-screen. Refer to the UUT Module –
Board and Connector Layout Drawing at the end of this section.
UUT Module Specifications
Up to four UUT Modules can be used with the Synergy Nano.
Each module can accept up to 16 thermocouple inputs for a total of 64 inputs.
The UUT module uses Type T Thermocouples.
RS-485 Data Communications are used to communicate with each UUT Module.
Hardware Connections
Disconnect power to the test chamber before starting. Lock-out / Tag-out your power source.
Follow the steps below to configure the UUT Module and to make necessary connections.
Once the items above have been removed, you will note the following cables that are connected to the UUT
Module.
Power Supply Cord: Black 2-wire cord with plug-in transformer, connected to Terminal Block P1.
RS-485 Interconnect 9-conductor ribbon cable, connected to Connector P6.
UUT Module Mechanical Drawing