Specifications
Synergy Nano Technical Manual, Revision -
Page 230
13.2 Setup Procedure
16 CHANNEL MODULE
ADDRESS SWITCH
SELECTION
MODULE NO SETTING
Module 1 1
Module 2 3
Module 3 5
1. Set up the UUT Module Address. Locate the small square
selector switch on the UUT Module labeled Address Switch. Turn
the dial on the switch to the proper Module Address setting
indicated in the table shown.
Module 4 7
2. Locate the RS-485 9-pin female plug, which is either plugged into the back of the RS-485 port on the
Data Communications panel, or is hanging freely behind the panel (for chambers not ordered with RS-
485).
Note: For chambers not initially supplied with an RS-485 port, look for the ribbon cable connected to the
RS-232 port. This ribbon cable is split and has a nine pin female RS-485 plug connected to one side of
the cable (and hanging freely).
3. Locate the RS-485 Interconnect 9-conductor ribbon cable connected to the UUT Module. There are four
additional 9-pin male connectors attached throughout the length of this rolled-up cable. These are
labeled M1 thru M4 on the UUT Module drawing. Connect one of the 9-pin male connectors from this
Interconnect ribbon cable to the RS-485 9-pin female connector located in Step 2.
4. Connect your T-Type Thermocouple input wires to the appropriate terminal blocks P2 thru P4 on the
UUT Module. Consult the Thermocouple Connection Tables on the UUT Module drawing for proper
terminations.
5. Be careful to route the thermocouple wires and the power supply cord safely through the appropriate
opening to avoid pinching.
BOARD PWR SUPPLY and
RS-485 CONNECTION TABLE
P1 CONNECTOR
DESIGNATION TERMINAL
Board GND. P1 - 1
Board POWER P1 - 2
COMM RX –
Note 1
P1 - 3
COMM TX + P1 - 4
COMM TX - P1 - 5
6. Plug in the UUT Module wall power transformer to a 120 Volt
outlet.
Verify that the green LED on the UUT Module is illuminated. It
should be on steadily.
Notes:
-The table at the right shows the board power supply connections
to the P1 connector.
-The additional RS – 485 terminals labeled COMM shown in the
table on the right are not required for normal wiring since the DB-
9 connector provides these connections.
COMM RX + P1 - 6
Address Selector
Switch