Specifications
Synergy Nano Technical Manual, Revision -
Page 231
7. Turn on power to your test chamber. Once the Synergy Nano completes the boot-up procedure, press
the COMM Navigation key. Open the RS-485 folder to arrive at the screen below.
Configure RS-485 Mode and Station Address to the values displayed in the screen below by pressing on
each item. Make the changes in the screens that follow. Configure the Number of UUTs using the chart
on the right.
Number of
Thermocouples
Number of UUTs
Value to be Entered
1 - 16 2
1 - 32 4
1 - 48 6
1 - 64 8
8. Cycle power to the chamber but leave the UUT power supply plugged in to the 120 V outlet.
After the Synergy Nano boots up successfully, verify that the green LED on the UUT Module is blinking.
This indicates that data communications have been established between the Synergy Nano and the UUT
Module.
Important Note: When you have more than one UUT Module, verify that the green LED blinks on each
module. Each module is queried sequentially as the Synergy Nano gathers the temperature data from
each of the modules.
9. Verify thermocouple temperature readings. Press the EVENTS Navigation key on the Synergy Nano.
Open the UUT Temperature folder to display the screen below
♦ Remember that each UUT Module can accept
up to 16 T/C inputs.
♦ Each screen displays 8 sensors labeled Sensor
1 thru Sensor 8. You can view all 16 inputs of a
module with 2 screens.
♦ Press the Next button to view the next set of
eight sensors for the first UUT Module. The
path display changes from UUT1 to UUT2. The
UUT2 screen will show the readings for actual
sensors No.’s 9 – 16.
♦ To view sensor inputs for UUT Modules. 2, 3
and 4, continue pressing the Next button. UUT
Module No. 2 data will be displayed by UUT3 /
UUT4, module by No. 3 UUT5 / UUT6, etc.
Important Note: If your screen shows “x.x” for sensor “Temp”, go back to Step 7 and verify the Number of
UUTs entered. X.x indicates that the module isn’t being monitored by the Synergy Nano.