The graph below shows the speed and control advantages of Synergy’s Cascade Control feature. Note that
the Air Temperature forces a faster product (UUT) temperature change by providing a greater delta t.
-100
-50
0
50
100
150
0 200 400 600 800 1000 1200 1400
Air Temp
UUT Temp
UUT setpoint
delta t=36 C
With
Cascade
delta t=16 C
Without
Cascade
14.1 Cascade Control Setup
14.1.1 Registering the Cascade Control feature
The cascade feature requires a registration key.
To register your cascade control feature, select
the cascade folder.
If your cascade control is not registered you will
encounter this screen. Press the Registration Key
text box to open the alphanumeric keypad. Enter
your registration number here. The cascade
registration number is available from Tidal
Engineering or your chamber manufacturer.