Specifications

Synergy Nano Technical Manual, Revision -
Page 238
14.2.4 Select Cascade Sensor
The Synergy Cascade control feature can use various input sensors to measure the product temperature.
The desired sensor is selected in the Settings screen shown below.
To select a product sensor to monitor the product
in Cascade, click on Sensor Select and press
the Change button.
Note: The Sensor Select field displays the
Sensor Select Number Code, which represents
the current sensor selected (211 in this example).
The code is explained below.
The sensor select screen shown below is used to
determine the code.
The Sensor Select screen displays two or three columns depending on the selected module. In the left most
column, you can specify the module; either the Olympic processor board or a UUT module.
Olympic Board Sensor
When you select Olympic board in the first
column, you can then select the sensor to
monitor.
Press the Accept button when finished.
The Sensor Select Number Code will appear in
the Settings screen shown above. This code is
derived from the Olympic Board Sensor Code
Chart shown below.
Olympic Board Sensor Code Chart
Sensor Value Description
RTD 1 110 Platinum 100 ohm RTD (Normally Chamber Air Temperature)
RTD 2 120 Platinum 100 ohm RTD (Normally not used in stand. chamber configurations)
Analog 1 130 0 to 5 Volt DC Signal (Normally Humidity sensor in T/H units)
Analog 2 140 0 to 5 Volt DC Signal (Normally not used in standard chamber configurations)
Analog 3 150 0 to 5 Volt DC Signal (Normally not used in standard chamber configurations)
Analog 4 160 0 to 5 Volt DC Signal (Normally not used in standard chamber configurations)