Specifications

Synergy Nano Technical Manual, Revision -
Page 243
15.0 SPACE AND ALTITUDE CHAMBERS
15.1 Space Chamber Setup
The Synergy Nano supports Space chamber applications also known as Thermal Vacuum chambers using
Granville Philips ION Vacuum gauges. In space chamber applications pressure is displayed and logged in
scientific notation. This section explains the setup procedure for Space Chamber applications and provides
examples. This section of the manual refers to the Synergy Nano Plus Olympic board wiring but the other
Synergy Nano configurations support this application as well.
CAUTION! : The Analog inputs on the controller will be damaged if the 10 V
Granville Philips output is attached without a voltage divider.
The setup consists of the following 5 steps:
1. Wire the Granville Philips Ion Vacuum pressure transducer to the controller.
2. Setup the chamber for the appropriate configuration; i.e. Generic Temperature /Pressure,
Generic Temperature/Humidity/Pressure, etc.
3. Setup the Hi Res input calibration for the pressure input.
4. Select the channel sensor for the pressure input.
5. Setup the Altitude calibration for the Granville Philips emission setting.