Specifications
Synergy Nano Technical Manual, Revision -
Page 244
Step 1. Wire the Granville Philips 10 Volt output thru a 2:1 voltage divider (1K/1K Ohm) as
shown below. Alternatively, Analog 2, 3 or 4 can be used. A precision 1% metal film resistor
suitable for this application is the Panasonic ERO-S2PHF1001 available from Digikey.
Signal Olympic Board Connector-Pin Reference
Analog Input 1 (See Note) P2-1 P2-7
Analog Input 2 P2-5 P2-7
Analog Input 3 P2-6 P2-7
Analog Input 4 P2-11 P2-7
Note: High Resolution Analog 1 should always be connected to a sensor or shorted to Analog
Common otherwise the open sensor detection circuitry (present on Analog 1 only) will
negatively affect the accuracy of the other inputs.
Step 2. Select the appropriate chamber setup
from the Chamber Setup folder and reboot as
instructed. For example, the Temperature
Pressure selection is shown at left.
Step 3. Setup the Hi Res input calibration for
the input used as shown at the left.
SETUP/Calibration/Input calibration/Hi
Res./Analog 1 (P2-1 to P2-7)
Set High Eng. Scale to 10 Volts
Set Low Eng. Scale to 0 Volts
Set High Volts Scale to 5 Volts
Set Low Volts Scale to 0 Volts