Specifications
Synergy Nano Technical Manual, Revision -
Page 248
The Pressure display on Channel 2 is in scientific notation as shown below. Note that the Pressure Set Point
has no effect on Thermal Vacuum chambers since pressure is not controlled.
Pressure is recorded in the log file in Torr in scientific notation.
To verify setup and calibration verify the values in the following table.
Output in Torr is 10^((Vhires*2)-n)
n, Altitude Setting V hi-res Volts DC Display
12 5.0 1.0e-2
11 5.0 1.0e-1
10 5.0 1.0e-0
12 0.0 1.0e-12
11 0.0 1.0e-11
10 0.0 1.0e-10
12 2.1 1.6e-8
11 2.1 1.6e-7
10 2.1 1.6e-6