Specifications
Synergy Nano Technical Manual, Revision -
Page 249
15.2 Altitude Chamber Setup
For conventional Altitude chambers (0-1000 Torr) the Synergy Controller can display the
pressure value in units of Torr or Kft.
The setup consists of the following 5 steps:
1. Connect pressure transducer to the controller.
2. Setup the chamber for the appropriate configuration; i.e. Generic Temperature /Pressure,
Generic Temperature/Humidity/Pressure, etc.
3. Setup the Hi Res input calibration for the pressure input. Adjust the Calibration\Input\High
Res\Analog1\ Folder and adjust to the appropriate High Engineering and Low Engineering
Scale for your transducer output in Torr.
4. Select the channel sensor for the pressure input.
5. Setup the Altitude calibration to 0.
To display pressure in Kft scroll to the bottom of the screen and set the Type to Torr-Kft as
shown below: