Specifications
Synergy Nano Technical Manual, Revision -
Page 317
Description Command Root Command
Usages
Command Syntax Range,
Units
Command
Example
Response Example
Setpoint 1 SP1 SP1 Set = SP1 X.X Range = R1L - R1H C / F = SP1 100.7 OK
SP1 Query ? SP1 Range = R1L - R1H C / F ? SP1 100.7
Setpoint 2 SP2 SP2 Set = SP2 X.X Range = R2L - R2H C / F / %RH = SP2 75 OK
SP2 Query ? SP2 Range = R2L - R2H C / F / %RH ? SP2 75.0
Setpoint 3 SP3 SP3 Set = SP3 X.X Range = R3L - R3H C / F / %RP = SP3 Off OK
SP3 Query ? SP3 Range = R3L - R3H C / F / %RP ? SP3 Off
Channel 1 Actual C1 C1 Query ? C1 Range = R1L - R1H C / F ? C1 25.0
Channel 2 Actual C2 C2 Query ? C2 Range = R2L - R2H C / F / %RH ? C2 50.0
Channel 3 Actual C3 C3 Query ? C3 Range = R3L - R3H C / F / %RP ? C3 Channel 3 is Off
Chamber On ON On Set = ON OK = ON OK
On Query ? On 1 if On, 0 if Off ? On 0
Chamber Off OFF OFF Set = OFF OK = OFF OK
Run Program RUN RUN Set = RUN OK = RUN OK
Run Query ? RUN Returns:
0 - Stop
1 - Run
2 - Pause
3 - Steady State
? RUN 0
Pause Program HOLD HOLD Set = HOLD OK = HOLD OK
Resume Program RSUM RSUM Set = RSUM OK = RSUM OK
Software Revision *IDN *IDN Query *IDN ? Returns Revision Info: Make, Model,
Serial #, Version
*IDN? Tidal Engineering, Synergy
Controller,Serial-02/0449,Version
2.6.8
Enable/Disable
UUT Monitoring
constants
UUT UUT Set = UUT ARG1 ARG2 ARG1 = UUT # (1 - 8)
ARG2 = 0/1 (Enable / Disable)
= UUT 1 1 OK
UUT Query ? UUT ARG1 Range: ARG1 = UUT # (1 - 8).
Response: 0/1 (Enabled / Disabled)
? UUT 1 0
UUT Readings UUTR UUTR Query ? UUT ARG1 Range: ARG1 = UUT # (1 - 8).
Response: comma delimited string
with 8 UUT temperature readings
If a sensor is not enabled, all values
returned will be 400.0 C or 752.0 F.
C / F
? UUTR 1 33.8,33.5,33.3,33.1,32.9,32.7,32.4,3
2.2