Specifications
Synergy Nano Technical Manual, Revision -
Page 325
Description Command Root Command
Usages
Command Syntax Range,
Units
Command
Example
Response Example
1CTY Query ? 1CTY 0 - CAP - Tube System
1 - Agree Logic
2 - Burn In Logic
3 - Standard XV Sys Logic
? 1CTY 0
2L1 2L1 2L1 Set = 2L1 ARG1 -100 - 100 % = 2L1 50 OK
2L1 Query ? 2L1 -100 - 100 % ? 2L1 50.00
2L2 2L2 2L2 Set = 2L2 ARG1 -100 - 100 % = 2L2 20 OK
2L2 Query ? 2L2 -100 - 100 % ? 2L2 50.00
2L3 2L3 2L3 Set = 2L3 ARG1 -100 to 100 C -148 to 212 F = 2L3 10 OK
2L3 Query ? 2L3 -100 to 100 C -148 to 212 F ? 2L3 50.00
2CTY 2CTY 2CTY Set = 2CTY ARG1 ARG1:
0 - CAP - Tube System
1 - Agree Logic
2 - Burn In Logic
3 - Standard XV Sys Logic
= 2CTY 1 OK
2CTY Query ? 2CTY 0 - CAP - Tube System
1 - Agree Logic
2 - Burn In Logic
3 - Standard XV Sys Logic
? 2CTY 0
L3 L3 L3 Set = L3 ARG1 0 - 100 % = L3 20 OK
L3 Query ? L3 0 - 100 % ? L3 20.00
L4 L4 L4 Set = L4 ARG1 0 - 100 % = L4 20 OK
L4 Query ? L4 0 - 100 % ? L4 20.00
L6 Query ? L6 -100 to 100 C -148 to 212 F ? L6 20.00
L6 L6 L6 Set = L6 ARG1 -100 to 100 C -148 to 212 F = L6 20 OK
L7 L7 L7 Set = L7 ARG1 0 - 100 % = L7 10 OK
L7 Query ? L7 0 - 100 % ? L7 10.00
L8 L8 L8 Set = L8 ARG1 0 - 100 % = L8 80 OK
L8 Query ? L8 0 - 100 % ? L8 80.00
L9 L9 L9 Set = L9 ARG1 -100 to 100 C -148 to 212 F = L9 50 OK
L9 Query ? L9 -100 to 100 C -148 to 212 F ? L9 50.00
L11 L11 L11 Set = L11 ARG1 0 - 100 % = L11 20 OK
L11 Query ? L11 0 - 100 % ? L11 20.00
L12 L12 L12 Set = L12 ARG1 0 - 100 % = L12 10 OK